Feasibility, acceptability and experiences of intrapartum fetal monitoring by caregivers using the Moyo device at Mbale regional referral hospital:
Busitema University
Background: Intrapartum fetal heart rate monitoring (FHRM) is essential for identifying fetal
distress and preventing asphyxia-related complications. However, inadequate FHRM due to
staffing shortages is a critical issue in low-resource settings such as Uganda, where intrapartumrelated
neonatal deaths are high. The Moyo device, a portable fetal heart rate (FHR) monitor,
reduced perinatal morbidity and mortality in a recently conducted trial in Mbale. However, its
potential remains underutilized due to a high patient-to-midwife ratio. Recent evidence suggests
that task-sharing FHRM with caregivers may address poor fetal monitoring practices, enabling
non-clinical individuals to support the timely identification of fetal distress.
Objective: This study assessed the feasibility, acceptability, and caregivers' experiences with
intrapartum FHRM using the Moyo device at Mbale Regional Referral Hospital (MRRH), Eastern
Uganda.
Methods: A pilot quasi-experimental study design was used employing a sequential explanatory
mixed-methods approach at MRRH. Caregivers of labouring women were consecutively enrolled,
trained, and guided to monitor FHR every 30 minutes using the Moyo device. Feasibility was
defined as achieving a mean monitoring interval within 30 ± 5 minutes, while acceptability was
assessed using a five-point Likert scale converted into percentage scores and categorized using
Bloom’s criteria. Experiences of caregivers who used the Moyo were explored through structured,
in-depth interviews. Quantitative data were analysed using Stata version 17, and qualitative data
were analysed thematically using Atlas. ti.
Results: A total of 75 caregivers participated, with a median age of 32 years (IQR: 25–42). Most
were female (n=70, 93.3%) and had primary-level education (n=35, 46.7%). The majority of
participants (n=74, 98.7%) were able to use the Moyo device to check and document FHR, with
65 (86.7%) reporting no technical issues. However, only 34.7% (n=26) of caregivers were able to
monitor FHR at 30-minute intervals, and 65.3% (n=49) of readings within 60 minutes. The median
number of FHR recordings was 3 (IQR: 1–10), and the mean duration of the active stage of labour
was 10.9 ± 9.7 hours. Feasibility was higher among caretakers who were first or second-degree
relatives (61.5%; p = 0.037). The majority of the caregivers (n=70, 93.3%) found FHRM using the
Moyo device acceptable. Caregivers reported positive experiences with the Moyo device, driven
by curiosity, reassurance about fetal safety, and a desire to learn.
Conclusion: Intrapartum FHRM by caregivers using the Moyo device was highly acceptable, and
most caregivers correctly used the device. However, adherence to the recommended 30-minute
monitoring interval remained low. Similar initiatives involving caregivers should prioritize
practical support systems like reminders and workflow integration to enable timely recordings and
improve adherence to recommended FHR monitoring.
Dissertation
Nakawuki, A.W. (2025). Feasibility, acceptability and experiences of intrapartum fetal monitoring by caregivers using the Moyo device at Mbale regional referral hospital: A pilot quasi experimental study. Busitema University. Unpublished dissertation.